Grounded Phoenix-to-Triax Cable Adapter 

PA-1004 Rev. A / June 2009 

 

 

 

Figure 3: MOSFET testing example using four Model 2600-FIX-TRX cable adapters 

 

 

For additional connection information, refer to the Series 2600A System SourceMeter

®

 Instruments Reference 

Manual (document number 2600AS-901-01). 

Handling and cleaning 

 
CAUTION

 

Performance to current leakage specifications depends upon a clean fixture. Avoid contamination 
that could cause degraded performance and increase current leakages beyond specified limits.  

 

 

 

Never use the Series 2600 chassis as a ground point for signal connections; it should only be used 
as a safety shield. Because high frequencies present on the chassis may result in higher noise, use 
the low-noise chassis ground banana jack instead. Use the chassis screw for connections to the 
Series 2600 chassis. 

 