Grounded Phoenix-to-Triax Cable Adapter
4
PA-1004 Rev. A / June 2009
Figure 3: MOSFET testing example using four Model 2600-FIX-TRX cable adapters
For additional connection information, refer to the Series 2600A System SourceMeter
®
Instruments Reference
Manual (document number 2600AS-901-01).
Handling and cleaning
CAUTION
Performance to current leakage specifications depends upon a clean fixture. Avoid contamination
that could cause degraded performance and increase current leakages beyond specified limits.
Never use the Series 2600 chassis as a ground point for signal connections; it should only be used
as a safety shield. Because high frequencies present on the chassis may result in higher noise, use
the low-noise chassis ground banana jack instead. Use the chassis screw for connections to the
Series 2600 chassis.