Example
:CAL:PROT:SAVE
Save calibration constants
:STEP
(:CALibration:PROTected:STEP<n>)
Purpose
To perform various calibration steps.
Format
:cal:prot:step<n>
Parameters
See Table B-2.
Description
The :CAL:PROT:STEP<n> command performs calibration at the various
points listed in Table B-2. See
Section 2
for details on test equipment and
connections.
Note
Calibration steps must be performed in the order listed in Table B-2, or an
error will occur.
Example
:CAL:PROT:STEP0 14
Perform cal step 0 (full-scale output).
Table B-2
Calibration step summary
Command
Description
:CALibration
Calibration subsystem.
:PROTected
Cal commands protected by password.
:STEP0 <nrf>
Output full-scale voltage (14V).
:STEP1 <nrf>
Calibrate output voltage setting using external DMM
reading.
:STEP2 <nrf>
Calibrate voltage measuring using external DMM reading.
:STEP3
Perform DVM input full-scale (14V) cal.
:STEP4 <nrf>
Output current (1.9A) for 5A full-scale cal.
:STEP5 <nrf>
Calibrate output current limit using calculated current.
:STEP6 <nrf>
Calibrate 5A measurement range using calculated current.
:STEP7
Output 5mA (500mA, Model 2303-PJ) nominal current.
:STEP8 <nrf>
Calibrate 5mA (500mA, Model 2303-PJ) measurement
range using calculated current.
深圳德标仪器
135-1095-0799
Содержание 2303
Страница 12: ...1 Performance Verification 135 1095 0799...
Страница 26: ...2 Calibration 135 1095 0799...
Страница 48: ...3 Disassembly 135 1095 0799...
Страница 53: ...3 6 Disassembly 135 1095 0799...
Страница 76: ...A Specifications 135 1095 0799...
Страница 81: ...A 7 Specifications 135 1095 0799...
Страница 82: ...B CalibrationReference 135 1095 0799...
Страница 90: ...C CalibrationPrograms 135 1095 0799...
Страница 97: ...D InstrumentControl 135 1095 0799...
Страница 104: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A 135 1095 0799...