PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL:
5mA to 5A, in 5mA steps.
TRIGGER DELAY:
0 to 100ms, in 10µs steps.
INTERNAL TRIGGER DELAY:
25µs.
HIGH/LOW/AVERAGE MODE:
Measurement Aperture Settings:
33.3µs to 833ms, in 33.3µs steps.
Average Readings:
1 to 100.
BURST MODE:
Measurement Aperture:
33.3µs.
Conversion Rate:
3600/second, typical.
Number of Samples:
1 to 5000.
Transfer Samples Across IEEE Bus in Binary Mode:
4800 bytes/second, typical.
LONG INTEGRATION MODE: Measurement Time
6
:
850ms (840ms) to 60 seconds in 16.7ms (20ms) steps.
GENERAL
ISOLATION (low - earth):
22VDC max.
PROGRAMMING:
IEEE-488.2 (SCPI).
USER-DEFINABLE POWER-UP STATES:
5.
REAR PANEL CONNECTOR:
8-position quick disconnect terminal block for output (4), sense (2), and DVM (2).
TEMPERATURE COEFFICIENT (outside 23°C ±5°C):
Derate accuracy specification by (0.1
×
specification)/°C.
OPERATING TEMPERATURE:
0° to 50°C (Derate to 70%).
0° to 35°C (Full power).
STORAGE TEMPERATURE:
–20° to 70°C.
HUMIDITY:
<80% @ 35°C non-condensing.
POWER CONSUMPTION:
150VA max.
REMOTE DISPLAY/KEYPAD OPTION:
Disables standard front panel.
DIMENSIONS:
89mm high
×
213mm wide
×
360mm deep (3
1
⁄
2
in
×
8
1
⁄
2
in
×
14
3
⁄
16
in).
NET WEIGHT:
3.2kg (7.1 lbs).
SHIPPING WEIGHT:
5.4kg (12 lbs).
INPUT POWER:
100–120VAC/220-240VAC, 50 or 60Hz (auto detected at power-up).
WARRANTY:
One year parts and labor on materials and workmanship.
EMC:
Conforms with European Union Directive directive 89/336/EEC
EN 55011, EN 50082-1, EN 61000-3-2 and 61000-3-3, FCC part 15 class B.
SAFETY:
Conforms with European Union Directive 73/23/EEC EN 61010-1,
UL 3111-1.
AC LINE LEAKAGE CURRENT:
450µA @ 110VAC, typ.; 600µA @ 220V, typ.
RELAY CONTROL JACK:
1-channel, sink 150mA max., 15V max. 5V output, 100mA max., also available on jack. Accepts 0.173 in Bantam-type
plug (CS-1003-1).
ACCESSORIES SUPPLIED:
User manual, service manual, output connector mating terminal (part no. CS-846).
ACCESSORIES AVAILABLE:
Model 2304-DISP:
Remote Display/Keypad (4.6 in
×
2.7 in
×
1.5 in). Includes 2.7m (9 ft) cable and rack mount kit.
Model 2303B:
2303 with blank front panel (only AC power indicator LED).
1
PLC = 1.00.
2
Following 15 minute warm-up, the change in output over 8 hours under ambient temperature, constant load, and line operating conditions.
3
Remote sense, at output terminals, 1000% load change; typical.
4
Remote sense, with 4.5m (15 ft) of 16 gauge wire and 1
Ω
resistance in each lead to simulate typical test environment, up to 1.5A load change.
5
Minimum current in constant current mode is 6mA.
6
60Hz (50Hz).
7
PLC = Power Line Cycle. 1PLC = 16.7ms for 60Hz operation, 20ms for 50Hz operation.
8
Display off.
9
Speed includes measurement and binary data transfer out of GPIB.
深圳德标仪器
135-1095-0799
Содержание 2303
Страница 12: ...1 Performance Verification 135 1095 0799...
Страница 26: ...2 Calibration 135 1095 0799...
Страница 48: ...3 Disassembly 135 1095 0799...
Страница 53: ...3 6 Disassembly 135 1095 0799...
Страница 76: ...A Specifications 135 1095 0799...
Страница 81: ...A 7 Specifications 135 1095 0799...
Страница 82: ...B CalibrationReference 135 1095 0799...
Страница 90: ...C CalibrationPrograms 135 1095 0799...
Страница 97: ...D InstrumentControl 135 1095 0799...
Страница 104: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A 135 1095 0799...