Configure and execute a 10-step linear list
sweep
The following example demonstrates the steps to set up the power
supply to output a linear list sweep from 1 V to 10 V in 1 V steps
(default settings). Each step in the list sweep will dwell for 3 seconds.
The list will be stored in list location 1. The output will turn on at 0 V
and return to 0 V at the end of the list execution.
To configure a list on the front panel:
1.
Press the
MEN U
key.
2.
Under Measure, highlight the
List
icon using either the
navigation control or the soft keys. Press
EN TER
.
3.
Select point 1 and press
EN TER
to put a check mark in the
box.
4.
Highlight
Insert
and press
EN TE R
to increment the points
display until it reads 10.
5.
Select the Voltage box in the point 1 row. Press
EN TER
.
6.
Make the value 1 V and press
E N TER
.
7.
Select the Current box in the point 1 row. Press
EN TER
.
8.
Make the value 2 A and press
E N TER
.
9.
Select the Dwell box in the point 1 row. Press
E N TER
.
10.
Make the value 3 s and press
EN TER
.
11.
Repeat steps 5 to 8 for points 2 to 10 and increase the voltage
by 1 V at each higher step.
12.
Select the
Save
button to store the list in memory location 1.
13.
Enter a name for the list on the displayed keypad and press
EN TER
.
14.
Press the
EXIT
key.
To delete a row on the front panel:
1.
Highlight the Points cell of the row you want to delete.
2.
Press
EN TER
to select the cell row.
3.
Navigate to
Delete
.
4.
Press
EN TER
.