Connections for testing
Before making the connections, prepare the wires as described in the
following table.
Usage
Specifications
Front panel binding posts
AWG 20 to AWG 12
Rear panel output terminals
AWG 20 to AWG 12
The wire must be heavy enough not to overheat while carrying
the short-circuit output current of the unit. Please meet the
wiring requirements described above.
Two-wire connection
The two-wire connections are used for basic operation when
maximum accuracy is not required. Keep the wire as short as possible
to reduce lead inductance and noise pickup. If you want to
compensate for the voltage drop in the load leads, use a four-wire
sense connection.
Two-wire (local sensing) DUT connection to front panel