2260B Series User Manual
78
Note
If the USB drive is not recognized, check to see
that the function settings for F-20 = 1 (page 90). If
not, reinsert the USB flash drive.
3.
Configure T-02 (Test sequnce
Load) to 1~10 (save memory slot)
T-02 range
1~10 (t001 ~t010)
4.
The test sequence will now be available in the
memory slot the test sequence was saved to.
Note
Error messages: If you load a file that is not
present on the USB drive “Err 002” will be
displayed on the display.
Run Test Sequence
Overview
A test sequence can be run from one of ten
memory slots.
Steps
1.
Before a test sequence can be run, it
must first be loaded into one of the
10 memory slots.
2.
Configure T-01 (Run Test
sequence) to 1~10 (save memory
slot)
T-01 range
1~10
3.
The test sequence will automatically start to
run.