OPERATION
69
2.
For each unit, hold the Function
key while turning the power on to
enter the power on configuration
settings.
3.
Configure F-93 (Master/Slave)
setting for each master/slave unit.
Unit
F-93
Master (local or series operation)
0
Slave unit (series)
4
4.
Cycle the power on the units (reset the power).
Note
Configuration settings can be checked for both the
master and slave units by pressing the Function
key.
Master-Slave
Operation
Only operate the power supplies in series if the
units are configured correctly.
1.
Turn on the master and slave unit. The slave
unit will only show the voltage of its own unit.
The master unit will show the combined
voltage of both units and the current.
Master unit
Slave unit
V
A
V