31
IEEE-488.2 common commands and queries
Table 4. IEEE-488.2 common commands and queries
Mnemonic
Name
Description
*CLS
*ESE <NRf>
*ESE?
*ESR?
*IDN?
*OPC
*OPC?
*OPT?
*RCL <NRf>
*RST
*SAV <NRf>
Clear status
Event enable command
Event enable query
Event status register
query
Identification query
Operation complete
command
Operation complete
query
Option identification
query
Recall command
Reset command
Save command
Clears all event registers
and Error Queue.
Program the Standard
Event Enable Register.
Read the Standard Event
Enable Register.
Read the Standard Event
Enable Register and clear
it.
Returns the manufac-
turer, model number,
serial number and firm-
ware revision levels of
the unit.
Set the Operation Com-
plete bit in the Standard
Event Status Register after
all pending commands
have been executed.
Places an ASCII “1” into
the output queue when
all pending selected
device operations have
been completed.
Returns an ID code that
indicates which memory
option is installed and
whether or not the
optional scanner card is
installed.
Returns the Model 2010
to the setup configura-
tion stored in the speci-
fied memory location.
Returned the Model
2010 to the *RST default
conditions.
Saves the current setup to
the specified memory
location.
Содержание 2010