9
Maximum inputs for the Model 2010 are summarized in
Table 1.
Basic measurement procedure
1. Connect the instrument as explained in “Instrument
connections”.
2. From the function keys, select the desired measure-
ment function.
3. Use the RANGE keys to select autoranging or a
manual range. Pressing the AUTO range key tog-
gles autoranging. You can select a different range
with the
▲
and
▼
RANGE keys.
4. Take a reading from the display.
Table 1. Maximum inputs
Function
Maximum input
DCV
1000V peak
ACV
750V rms, 1000V peak, 8
×
10
7
V•Hz
DCI
3A, 250V dc
ACI
3A 250V rms
FREQ (PERIOD)
1000V peak, 8
×
10
7
V•Hz