1-26
Return to
Section Topics
2000-903-01 Rev. C / October 2007
Section 1: Introduction
Model 2000 Digital Multimeter Quick Start Guide
:TEMPerature
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:STATe <b>
:STATe?
:TCouple
:TYPE <name>
:TYPE?
:RJUNction
:RSELect <name>
:RSELect?
:SIMulated <n>
:SIMulated?
:REAL
:TCOefficient <n>
:TCOefficient?
:OFFSET <n>
:OFFSet?
Select filter type: (MOVing or
REPeat).
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
Path to configure
thermocouple:
Select thermocouple type (J,
K, or T).
Query thermocouple type.
Path to configure reference
junction:
Select reference type
(SIMulated or REAL).
Query reference type.
Specify simulated
temperature in °C (0 to 50).
Query simulated
temperature.
Path to configure real
reference junction:
Specify temp coefficient
(-0.09999 to 0.09999).
Query temp coefficient.
Specify voltage offset at 0°C
(-0.09999 to 0.09999).
Query voltage offset.
:FREQuency
:APERture
:APERture?
:THReshold
:VOLTage
:RANGe <n>
:RANGe?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
Path to configure frequency.
Sets gate time for period
measurements (0.01-1.0s).
Query period gate time.
Path to select the threshold
voltage range:
Select threshold range (0 to
1010).
Query threshold range.
Specify reference (0 to
1.5e7)
Enable or disable reference.
Query state of reference.
Use input signal as
reference.
Query reference value.
Specify measurement
resolution (4 to 7).
Query resolution.
Table 1-10
SENSe command summary (cont.)
Command
Description