1-20
Return to
Section Topics
2000-903-01 Rev. C / October 2007
Section 1: Introduction
Model 2000 Digital Multimeter Quick Start Guide
Table 1-9
ROUTe command summary
Command
Description
:ROUTe
:CLOSe <chan
num>
:STATe?
:OPEN:ALL
:MULTiple
:CLOSe <list>
:STATe?
:OPEN <list>
:SCAN
[:INTernal] <list>
[:INTernal]?
:EXTernal <list>
:EXTernal?
:LSELect <name>
:LSELect?
Commands to control scanner
card:
Close specified channel (1 to
10) or channel pair (1 to 5).
Query closed channel (or
channel pair).
Open all input channels
(1 through 10).
Path to close and open
multiple channels:
Close specified channels (1 to
11).
Query closed channel.
Open specified channels (1 to
11).
Path to scan channels.
Specify internal scan list (2 to
10 channels).
Query internal scan list.
Specify external scan list (2 to
800 channels).
Query external scan list.
Select scan operation
(INTernal, EXTernal, or
NONE).
Query scan operation.
Table 1-10
SENSe command summary
Command
Description
[:SENSe[1]]
:FUNCtion <name>
:FUNCtion?
:DATA?
:FRESh?
:HOLD
:WINDow <NRf>
:WINDow?
:COUNt <NRf>
:COUNt?
:STATe <NRf>
:STATe?
Select measurement
function: ‘VOLTage:AC’,
VOLTage :DC’, RESistance’,
‘FRESistance’,
‘CURRent:AC’, ‘CURRent:
DC’ ,
‘FREQuency’,‘TEMPerature’,
‘PERiod’, ‘DIODe’,
“CONTinuity’.
Query function.
Return the last instrument
reading.
Returns a new (fresh)
reading.
Path to control Hold feature:
Set Hold window (%); 0.01 to
20.
Query Hold window.
Set Hold count; 2 to 100.
Query Hold count.
Enable or disable Hold.
Query state of Hold.