Operation
3-3
Scanner card detection
Power-up detection
The scanner card is detected only at power-on. If the card is plugged in after the power is
turned on, the multimeter will not recognize the presence of the card.
CAUTION
Plugging in the scanner card with power turned on may result in damage to
both the Model 2000-SCAN-20 and the multimeter.
If the card is not present at power-on, scanner bus commands or queries will generate a
“Missing hardware error,” and front panel operations pertaining to the scanner will inform you
that no scanner is present.
Scanner option bus query
*OPT? is an IEEE 488.2 common query that determines whether or not a scanner card is
installed. Also, the Model 2000-20 multimeter can distinguish between a Model-2000-SCAN
10-channel card and the Model-2000-SCAN-20 20-channel card. The response to this query is
as follows:
•
0: Scanner not installed.
•
0,10CH: Model 2000-SCAN 10-channel card installed.
•
0,20CH: Model 2000-SCAN-20 20-channel card installed.
Refer to the multimeter manual for more details on using the *OPT? query.
Содержание 2000-SCAN-20
Страница 11: ......
Страница 12: ...1 General Information General Information ...
Страница 17: ...1 6 General Information ...
Страница 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Страница 35: ...2 18 Card Connections and Installation ...
Страница 36: ...3 Operation Operation ...
Страница 51: ...3 16 Operation ...
Страница 52: ...4 Service Information Service Information ...
Страница 63: ...4 12 Service Information ...
Страница 64: ...5 Replaceable Parts Replace ableParts ...
Страница 67: ...5 4 Replaceable Parts ...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Страница 79: ......
Страница 80: ......