2-8
Card Connections and Installation
Typical connecting schemes
The following paragraphs discuss typical connections for the scanner card.
Voltage connections
shows typical connections for voltage measurements. Note that all channels are
used in the 2-pole mode and that up to 20 voltage sources can be switched with this
configuration. This basic configuration can be used for the following types of measurements:
•
DCV
•
ACV
•
Frequency (voltage only)
All channels (1 through 20) can be used with this configuration.
Figure 2-4
Connections for voltage scanning
CH 1
CH 20
V
HI
LO
V
HI
LO
To DMM
Input
HI
LO
2000-SCAN-20 Card
Voltages
Under Test
HI
LO
HI
LO
OUT A
OUT B
HI
LO
HI
LO
J1036
J1035