Kegel Toner Скачать руководство пользователя страница 25

24 

 

the metal plates on the 

Kegel 

Trainer

 and your skin. 

that in a few weeks of stimulation the contact 
improves. If it does not, this unit may not be 
suitable  for  you.  You  may  need  to  contact 
your  healthcare  professional  to  discuss 
other suitable options. 

iii) The 

Kegel Trainer

 supplied with the unit has 

a 28 mm diameter. An optional 32 mm 

Kegel 

Trainer

, part no. 

X-VPM

, is available. 

No sensation 
and no LEADS 
alarm showing 

Intensity  level  is  not  high 
enough 

and/or 

reduced 

sensitivity  in  the  area  being 
treated. 

i)  Please  make  sure  you  are  increasing  the 

intensity high enough. Most people will start 
to feel the stimulation in their hand at around 
25  mA and with the 

Kegel Trainer

 inserted 

you will need to increase the intensity higher 
to around 40 mA 

– 60 mA. Max power is 99 

mA.  Everyone  is  different  so  just  keep 
increasing  the  intensity  until  you  can feel  it. 
The intensity increases in very small steps of 
1 mA. 

ii) You  may  have  reduced  sensitivity  due  to 

previously 

damaged 

or 

desensitized 

pudendal  nerves  (this  can  happen  in 
childbirth  or  some  surgical  procedures). 
Please consult your healthcare professional. 

No sensation 
on one side of 
the Kegel 
Trainer (or 
electrode) 

Position  is  not  optimal 

– 

needs adjusting. 

The  current  flows  from  one  side  of  the 

Kegel 

Trainer

 to the other, so it is not possible to have 

one side “not working”. However, the strength 
of the sensation depends on how close to the 
nerve  the  current  flows,  and  also  in  which 
direction it flows relative to the nerve. You can 
try slightly adjusting the position on the 

Kegel 

Trainer

,  or  exchanging  the  connection  of  the 

wires in the 

Kegel Trainer

Sudden 
change in 
sensation 

If  you  disconnect  and  re-
connect  a  few  minutes  later, 
the  signal  will  feel  quite  a  lot 
stronger.  

Always 

return 

strength 

to 

zero 

after 

disconnecting the lead or the 

Kegel Trainer

The patient  is an  intended  operator. There  are  no user-serviceable  parts  inside the 
unit, and no calibration is required.  

 

If  the  above  review  has  failed  to  resolve  your  problem,  or  to  report  unexpected 
operation or events, call TensCare or your local dealer (address on back cover) for 
advice.    

 

Contact TensCare customer service on +44 (0) 1372 723 434. Our staff are trained to 
assist you with most issues you may have experienced, without the need to send your 
product in for service. 

Содержание Toner

Страница 1: ...0 Kegel Toner Pelvic Floor Trainer INSTRUCTIONS FOR USE READ CAREFULLY BEFORE USE...

Страница 2: ...sert the Kegel Trainer Press and hold the ON button to switch the device on Regulate the output intensity with the buttons and Press and hold the OFF button to switch the device off ON Button P UP DOW...

Страница 3: ...y tested products for the applications in the areas of gentle electrotherapy muscle toning continence management and pain relief during labour Please read these instructions for use carefully and keep...

Страница 4: ...PRESET PROGRAMS 12 8 CONTENT 12 9 UNIT INFORMATION 13 9 1 CONTROLS DISPLAY 13 9 2 OPERATING INSTRUCTIONS 14 10 SETTING UP AND USING THE KEGEL TONER 16 10 1 INSTALLATION OF BATTERIES 16 10 2 CONNECTIN...

Страница 5: ...ue number so that the device can be identified Atmospheric Pressure indicates the atmospheric limits to which the medical device can be safely exposed Date of Manufacture indicates the date which the...

Страница 6: ...he Kegel Toner is intended to provide electrical stimulation and neuromuscular re education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress incontinence in w...

Страница 7: ...the front to the end of the spinal column or tail bone in the back The pelvic floor muscles Assist in supporting the abdominal and pelvic organs Work with the abdominal and back muscles to stabilize...

Страница 8: ...d increase your sexual pleasure A continence therapist can help you learn how to exercise your pelvic floor Doing just a few pelvic floor exercises every day will help to treat bladder weakness or pro...

Страница 9: ...on In rehabilitation EMS is a well established method for treatment of a broad field of musculoskeletal diagnoses as well as pelvic floor weakness Electrical stimulation of an intact peripheral nervou...

Страница 10: ...on is used when failure to follow the instructions may result in a minor or moderate injury or damage to the device or other property Notes are used to provide clarification or recommendation CONTRAIN...

Страница 11: ...ment to prevent injury Caution Kegel Toner is intended for indoor use only because it is an electronic device Caution Only use recommended cleaning solutions that leave no residue to prevent irritatio...

Страница 12: ...steel in the Kegel Trainer s metal plates contain some Nickel This could cause a reaction if you have a Nickel allergy Use with caution if you have a copper IUD If discomfort occurs discontinue treatm...

Страница 13: ...how results Successful treatment requires use once a day for one to three months You may notice an improvement after about four weeks It helps to keep a diary of leakage problems so that you can measu...

Страница 14: ...RMATION 9 1 CONTROLS DISPLAY Program selected Program selector Time selector OFF button ON button Keypad Lock Program duration min Strength selected mA Strength up Strength down Low Battery Output act...

Страница 15: ...ton press STRENGTH CONTROLS The buttons marked and are the strength controls To increase strength press and hold down the button until required strength is achieved To decrease the strength press and...

Страница 16: ...releasing the T button you can manually override and select a shorter treatment time of 10 minutes You may wish to use this during your early training sessions The length of time that you have select...

Страница 17: ...GEL TONER 10 1 INSTALLATION OF BATTERIES 1 Remove battery cover by pulling on tag 2 Insert batteries Ensure that the batteries are inserted the right way as shown in battery compartment and that the r...

Страница 18: ...ghly with lots of water Keep batteries out of the reach of small children Caution NEVER attempt to recharge an alkaline battery Risk of explosion Caution Do not mix old new or different types of batte...

Страница 19: ...and should be in contact with the main part of the muscle at all times The tissues close to the entrance are more sensitive so you should avoid stimulating them Note The flange should not be inserted...

Страница 20: ...all this depends very much on the individual and any pre existing physical conditions so slowly increase the intensity by repeatedly pressing the button until you begin to feel the muscles around you...

Страница 21: ...TS At the end of its life please dispose of your Kegel Toner responsibly Anything electrical or electronic should not be treated as domestic waste and simply thrown away About half of the states curre...

Страница 22: ...struction booklet enclosed or if product repairs are carried out without authority from TensCare Ltd We will repair or at our option replace free of charge any parts necessary to correct material or w...

Страница 23: ...eipt For hygiene reasons please do not include used probe trainer Send only the unit and the lead wire Please return the unit and warranty card see page 29 TensCare Ltd PainAway House 9 Blenheim Road...

Страница 24: ...prevent anyone from increasing the intensity to a high level without firm contact between the machine and the skin A connection error can occur if 1 A break has developed within one of the two lead w...

Страница 25: ...e surgical procedures Please consult your healthcare professional No sensation on one side of the Kegel Trainer or electrode Position is not optimal needs adjusting The current flows from one side of...

Страница 26: ...o 93 RH non condensing Atmospheric pressure 700hPa to 1060hPa Temperature range 25 70 C Humidity Up to 93 RH non condensing Atmospheric pressure 700hPa to 1060hPa TYPE BF EQUIPMENT Equipment providing...

Страница 27: ...r base stations walkie talkies can affect this equipment and should be kept at least a distance d 3 3 m away from the equipment Note As indicated in Table 6 of IEC 60601 1 2 2007 for ME EQUIPMENT a ty...

Страница 28: ...WARRANTY NAME ADDRESS POSTCODE DAYTIME TELEPHONE E MAIL MODEL DATE OF PURCHASE ATTACH PROOF OF PURCHASE DO NOT SEND IN KEGEL TRAINER RETAILERS NAME RETAILERS ADDRESS RETAILERS POSTCODE BRIEF DESCRIPT...

Страница 29: ...28 Notes...

Страница 30: ...29...

Страница 31: ...e limited https plus google com TenscareLtdEpsom https uk pinterest com TensCareLtd https tenscareltd wordpress com https www youtube com channel UCzpik9dmLlJ3j0aHOpQ 0sg EC Declaration of Conformity...

Страница 32: ...31 Distributed by Manufactured by TensCare Ltd 9 Blenheim Road Epsom Surrey KT19 9BE UK Tel 44 0 1372 723434 www tenscare co uk Pub No I KT UK Version 1 0 04 19...

Отзывы: