Pulsair Intellipuff Tonometer by KEELER
• To prevent condensation from forming, allow instrument to come to room temperature
before use.
• Only mount on wall according to Keeler Instructions.
• This product should be used in a room with low / dimmed lighting.
• Before using the Pulsair IntelliPuff Tonometer, press the Demo button for 1 second
to dispel any minute particles of dust or moisture which may have settled whilst the
instrument was not in use.
• For indoor use only (protect from moisture).
• There are no user serviceable parts inside. Contact authorised service representative for
further information.
• Follow guidance on cleaning / routine maintenance to prevent personal injury / damage to
equipment.
• Failure to carry out recommended routine maintenance as per the instructions in this IFU
may reduce the operational lifetime of the product.
• At product end of life dispose of in accordance with local environmental guidelines (WEEE).
2.3 CONTRAINDICATION
There is no restriction to patient population this device can be used with other than those
outlined in the contraindications stated below.
Accuracy of IOP measurements is known to be affected by variations and changes in corneal
rigidity due to differences in corneal thickness, intrinsic structural factors or corneal refractive
surgery. It is recommended that these factors are considered during IOP measurement.
3. CLEANING INSTRUCTIONS
3.1 CLEAN THE PUFF TUBE LENS ON A WEEKLY BASIS:
1. Moisten a cotton bud with Isopropyl Alcohol.
2. Move the tip of the bud around the lens in a
circular motion.
3. After one circle the bud should be discarded
to avoid smearing on the lens.
4. Look at the Puff Tube lens from the patient’s
side, if traces of tear film can still be seen,
repeat above steps until clear.
Note:
Care should be taken not to damage the Puff
Tube assembly during cleaning.
CAUTION: Never use a dry cotton bud or tissue to clean the Puff Tube lens.
Never use a silicone impregnated cloth or tissue to clean the Puff Tube
lens.
Содержание Pulsair Intellipuf
Страница 1: ...Non Contact Tonometer INSTRUCTIONS FOR USE EN BG RU HY HR SR...
Страница 2: ......
Страница 4: ......
Страница 28: ...24 BG Pulsair Intellipuff KEELER 1 Pulsair IntelliPuff 1 1 1 2 Pulsair IntelliPuff 2 2 1 Keeler 24 Keeler Ltd...
Страница 29: ...25 BG Pulsair Intellipuff KEELER 2 2 Keeler Ltd Keeler Keeler EP29 32777 Pulsair IntelliPuff Keeler...
Страница 30: ...26 BG Pulsair Intellipuff KEELER Keeler Pulsair IntelliPuff Demo 1 2 3 3 3 1 1 2 3 4...
Страница 32: ...28 BG Pulsair Intellipuff KEELER 4 Pulsair IntelliPuff 1 mmHg 7 1 Pulsair IntelliPuff LED 2 3 LED...
Страница 33: ...29 BG Pulsair Intellipuff KEELER 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Страница 70: ...66 HY KEELER Pulsair Intellipuff 1 Pulsair IntelliPuff 1 1 1 2 Pulsair IntelliPuff 2 2 1 Keeler 24 Keeler Ltd...
Страница 71: ...67 HY KEELER Pulsair Intellipuff 2 2 Keeler Ltd Keeler Keeler EP29 32777 Pulsair IntelliPuff Keeler...
Страница 72: ...68 HY KEELER Pulsair Intellipuff Keeler Pulsair IntelliPuff 1 Demo WEEE 2 3 3 3 1 1 2 3 4...
Страница 74: ...70 HY KEELER Pulsair Intellipuff 4 Pulsair IntelliPuff 1 7 1 On Off Pulsair IntelliPuff On Off LED 2 3 LED...
Страница 75: ...71 HY KEELER Pulsair Intellipuff 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Страница 92: ...88 KEELER Pulsair IntelliPuff RU Keeler Keeler EP29 32777 Pulsair IntelliPuff Keeler...
Страница 93: ...89 KEELER Pulsair IntelliPuff RU Keeler Pulsair IntelliPuff Demo 1 WEEE 2 3 3 3 1 1 2 3 4...
Страница 95: ...91 KEELER Pulsair IntelliPuff RU Pulsair Desktop 1 7 1 On Off 2 3 Print...
Страница 96: ...92 KEELER Pulsair IntelliPuff RU 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Страница 101: ...97 KEELER Pulsair IntelliPuff RU 9 STBY WAIT OK OK OD 14 OD OS 0 1 0 1 1 25 25 25 run test 15 OD OS...
Страница 131: ......