Table 3.1 below provides the IOP correction values.
Table 3.1 - IOP correction values
Corneal thickness (micrometers)
Correction values (mmHg)
405
7
425
6
445
5
465
4
485
3
505
2
525
1
545
0
565
-1
585
-2
605
-3
625
-4
645
-5
665
-6
685
-7
705
-8
Correction values according to corneal thickness of 545 micrometers. These correction values are modified from
the work of Doughty and Zamen. This chart was reproduced from the Review of Ophthalmology, July 2002. Leon
Herndon, MD, Duke University, Glaucoma Service, pages 88, 89, 90.