Next
Home
Back
6
4. Cleaning instructions
Only manual non-immersion cleaning as described should
be used for this instrument. Do not autoclave or immerse in
cleaning fluids.
a Wipe the external surface with a clean absorbent, non-
shedding cloth dampened with a water / detergent solution
(2% detergent by volume) or water / isopropyl alcohol
solution (70% IPA by volume). Avoid optical surfaces.
b Ensure that excess solution does not enter the instrument.
Use caution to ensure cloth is not saturated with solution.
c Surfaces must be carefully hand-dried using a clean non-
shedding cloth.
d Safely dispose of used cleaning materials.
4.1 Sterilisation
Keeler recommend that ear speculum are used once.
However, the ear speculum maybe sterilised at 134° for a dwell
time of 10 minutes in a steam steriliser.
Single use - repeated use could cause infection.