17
EN
D-KAT Z-Type Tonometer by KEELER
Cause:
The rings are too far to the right.
Correction:
Using the joystick, move the slit lamp to
the right.
Cause:
The reading in this position is considerably
higher than real ocular pressure.
Correction:
Using the slit lamp height adjustment
mechanism, lower the slit lamp until the two
fluorescein semi circles are equal in size.
The measurement pressure will then be
reduced.
8.4 THE INSIDE MARGINS OF THE FLUORESCEIN RINGS ARE NOT ALIGNED
AND TOUCHING
Cause:
The semicircular images are well centred.
The outer margins are aligned but the inner
margins are not.
Correction:
Increase pressure by rotating the
measurement drum.
Cause:
In this case, the inner margins of one
semicircle are aligned with the outer
margins of the other.
Correction:
Increase pressure by rotating the
measurement drum.