WARRANTY 

The meter is warranted to be free for one year from the date of purchase. (Electrode and battery are not included) 

This warranty covers normal operation and does not cover battery, misuse, abuse, alteration, tampering, neglect, 

improper maintenance or damage resulting from leaking batteries. Proof of purchase is required for warranty repairs. 

Warranty is void if the meter has been opened. 

NOTE 

1.

 

Please operate the meter Under the guidance of professionals. 

2.

 

Soak the electrode into the distilled water or purified water for 5 – 30 minutes to activate before use. 

3.

 

Please use the standard buffer solutions for pH calibration, and operate it by the calibration steps shown above. 

DO NOT press CAL key in NON calibration mode, it may cause the calibration Error. 

4.

 

The pH buffer powder should be fully dissolved in water (250ml). 

5.

 

In calibration mode, the electrode should be submerged in the buffer solution. Especially pay attention to the 

Electrode fixed collar if it is loosen when calibration, it may cause the buffer solution enter the inside of the 

meter. 

How to make pH buffer solution 

There are 3 pH buffer powder supplied with the meter

 when you buy them, each powder 

should be Fully dissolved in 

water(250ml). DO NOT use the metal containers. 

Take the formulation of pH 6.8 buffer solution for example:   

1. Put the pH 6.86 buffer material (in the green sachet) in a jar (the volume should be no less than 250ml).   

2. Fill the jar with 250ml distill water. 

3. Place a glass stick in the solution, swirling it until the white powder dissolved. 

4.    Store the buffer solution in the cool and dry place. Attach a label on it for further use. 

5. For the further calibration, Ensure that there is a constant supply of fresh buffer solution in contact with the 

probe. Discard the solution after use. 

The way of the formulation of other buffer is same like that . Please mark it clear and prepare to use it for 

the future. 

   

The use of standard buffer

 

The buffer solutions (250ml) can be used several times by divided into portions which should be kept in a cool and 

dry place(20

-25

). Never reuse the used portion. The electrode should be fully covered by buffer solution when 

calibration. 

Содержание CT-6821

Страница 1: ... off the protection cap Any damage may cause the invalid of the electrode The duration period of the glass electrode is one year from the date of purchase whether you use it or not please change the electrode in time The electrode can not be used under the dry condition always soak the electrode into the distilled water or purified water for 5 30 minutes to activate before use Figure Part Name 1 B...

Страница 2: ... sensor ATC auto temperature compensation Interchangeable Three point calibration for pH 4 0 pH 6 8 and pH 9 1 Already calibrated can be used directly SPECIFICATIONS Display LCD size 20 mm x 27 mm Consumption Measurement Range pH 0 0 to 14 0 pH ORP 500mV Resolution pH 0 1 pH ORP 2mV Accuracy pH 0 1 pH After calibration ORP 5mV pH Calibration pH4 0 pH6 8 pH9 1 Three Point Calibration ORP Calibratio...

Страница 3: ...screen 2 Press HOLD key to select ORP on the screen 3 Then press CAL to confirm and the meter is ready to test in ORP mode Backlight Mode 1 Press the ON OFF key for 3 seconds until the pH display on the screen 2 Press HOLD key to select LED on the screen 3 Then press CAL to confirm the backlight mode 4 Power off the meter to close the backlight mode Hold Feature Press HOLD to freeze the current re...

Страница 4: ...The pH value will be updated until a stable reading is reached The data maybe not accurate if the electrode is not be activated 4 press CAL key for 3 seconds until the text CAL appear on the display and release the key 5 Text SA will showing in the LCD denote that the current calibration result is saved 6 The text END will be displayed at the end The meter is ready for pH measurement Overrage Remi...

Страница 5: ...fer solution Especially pay attention to the Electrode fixed collar if it is loosen when calibration it may cause the buffer solution enter the inside of the meter How to make pH buffer solution There are 3 pH buffer powder supplied with the meter when you buy them each powder should be Fully dissolved in water 250ml DO NOT use the metal containers Take the formulation of pH 6 8 buffer solution fo...

Отзывы: