  Pen Meter pH & ORP Users Manual 

                              

CT-6821 

 Caution 

 

Please read the manual carefully before using the meter. 

 

The glass electrode at the bottom of the pen is fragile, please use it carefully after taking off the protection cap. 

Any damage may cause the invalid of the electrode. 

 

The duration period of the glass electrode is one year from the date of purchase whether you use it or not, 

please change the electrode in time. 

 

The electrode can not be used under the dry condition, always soak the electrode into the distilled water or 

purified water for 5 – 30 minutes to activate before use.   

 

 

                                                                     

 

 

 

 

 

 

 

 

Figure 

Part Name 

Battery Storage Section   

pH& ORP Display 

Temperature Display 

CAL Calibration Key 

HOLD key 

Power ON/OF Key 

Electrode fixed collar 

Electrode Protection Cap 

Содержание CT-6821

Страница 1: ... off the protection cap Any damage may cause the invalid of the electrode The duration period of the glass electrode is one year from the date of purchase whether you use it or not please change the electrode in time The electrode can not be used under the dry condition always soak the electrode into the distilled water or purified water for 5 30 minutes to activate before use Figure Part Name 1 B...

Страница 2: ... sensor ATC auto temperature compensation Interchangeable Three point calibration for pH 4 0 pH 6 8 and pH 9 1 Already calibrated can be used directly SPECIFICATIONS Display LCD size 20 mm x 27 mm Consumption Measurement Range pH 0 0 to 14 0 pH ORP 500mV Resolution pH 0 1 pH ORP 2mV Accuracy pH 0 1 pH After calibration ORP 5mV pH Calibration pH4 0 pH6 8 pH9 1 Three Point Calibration ORP Calibratio...

Страница 3: ...screen 2 Press HOLD key to select ORP on the screen 3 Then press CAL to confirm and the meter is ready to test in ORP mode Backlight Mode 1 Press the ON OFF key for 3 seconds until the pH display on the screen 2 Press HOLD key to select LED on the screen 3 Then press CAL to confirm the backlight mode 4 Power off the meter to close the backlight mode Hold Feature Press HOLD to freeze the current re...

Страница 4: ...The pH value will be updated until a stable reading is reached The data maybe not accurate if the electrode is not be activated 4 press CAL key for 3 seconds until the text CAL appear on the display and release the key 5 Text SA will showing in the LCD denote that the current calibration result is saved 6 The text END will be displayed at the end The meter is ready for pH measurement Overrage Remi...

Страница 5: ...fer solution Especially pay attention to the Electrode fixed collar if it is loosen when calibration it may cause the buffer solution enter the inside of the meter How to make pH buffer solution There are 3 pH buffer powder supplied with the meter when you buy them each powder should be Fully dissolved in water 250ml DO NOT use the metal containers Take the formulation of pH 6 8 buffer solution fo...