First Step

®

 Plus™ Operations Manual

4-1 

Care and Cleaning

®

4

CARE AND CLEANING

The following are the KCI recommended daily and weekly cleaning and infection control procedures for the First 
Step

®

 

Plus

 Therapeutic Mattress Support System while it is in use. If this product is to be purchased, or utilized 

for long term rentals, it is recommended that KCI be contacted for recommended Infection Control Procedures to 
be utilized by the facility.

It is recommended that all chapters of this manual be reviewed prior to product use. Carefully read the

 

Contraindications, Risks and Precautions and Safety Tips 

section in the 

Introduction 

chapter of this manual 

prior to performing care and cleaning procedures on the First Step

®

 

Plus

 Therapeutic Mattress Support System.

Follow all regulations outlined in OSHA Occupational Exposure to Bloodborne Pathogens 29 CFR 1910.1030 
regarding Infection Control Procedures to be used when cleaning First Step

®

 

Plus

.

DAILY CARE AND CLEANING OF FIRST STEP

®

 PLUS™ PRODUCT 

WHILE IN USE

NOTE:

The patient need not be removed from First Step

®

 

Plus

 product when performing daily cleaning 

procedures. The patient may be bathed as patient support surface is being cleaned.

1.

If Side Rails are used (refer to 

Risks and Precautions

 and 

Safety Tips

 in the 

Introduction

 Chapter of this 

manual). Verify Side Rails on both sides of bed are up and locked.

2.

Inflate surface fully by adjusting all three air flow control knobs to their highest setting.

3.

If Side Rails are used, lower Side Rail only on Caregiver's side of bed.

4.

Log roll patient to opposite side of surface.

5.

Clean patient as required.

NOTE:

Avoid spilling fluids on any part of the air supply unit. Fluids remaining on the electronic controls can 
cause corrosion which may cause the electronic components to fail. Component failures may cause 
the air supply unit to operate erratically possibly producing a potential hazard to patient or staff.

6.

Wipe Cover Sheet as required with soap and water or approved hospital germicide. 

7.

Move cover sheet for access and wipe soiled mattress overlay as required. 

Replace and launder extensively 

soiled cover sheet or mattress overlay.

8.

Dry surface and cover sheet with towel. Remaining moisture will evaporate rapidly with air flow.

9.

Move cover sheet back into position.

10.

Move patient to clean, dry side of surface and repeat Steps 4 through 11 on the opposite side.

11.

Return patient to center of surface.

12.

Return surface to previous air pressure settings.

WEEKLY CARE AND CLEANING OF FIRST STEP

®

 PLUS™ 

PRODUCT WHILE IN USE

In addition to the daily care and cleaning described above, the First Step

®

 

Plus

 cover sheet should be replaced 

weekly or more often as needed. Excessively soiled mattress overlays and cover sheets should be replaced and 
laundered as described in the 

Infection Control

 section of the 

Preparation for Use

 chapter of the 

First Step

®

 

Plus

 

Maintenance

 

Manual

.

Rev. B 05/05

Содержание FIRST STEP Plus

Страница 1: ...KCI USA Inc San Antonio Texas 78265 9508 Call Toll Free 1 800 275 4524 Fax www kci1 com P O Box 659508 1 800 275 3417 R TM Plus ...

Страница 2: ......

Страница 3: ... out only by qualified personnel authorized by KCI The electrical installation of the room must comply with the appropriate electrical wiring standards The product must be used in accordance with this manual this product s Maintenance Manual and Quick Reference Guide and all applicable labeling The only servicing to be carried out by the user is described in the Maintenance Chapter of the First St...

Страница 4: ...First Step Plus Operations Manual ...

Страница 5: ...irst Step Plus Product 3 1 Patient Transfer From First Step Plus Product 3 1 Ambulatory Patient Transfer 3 1 Non Ambulatory Patient Transfer 3 1 Patient Transport 3 2 Patient Bathing 3 2 Bedpan Placement 3 3 Bedpan Removal 3 3 Repositioning Patient Using Dri flo Underpads 3 5 Incontinence Drainage 3 5 Skin Care 3 5 General Operation 3 5 CARE AND CLEANING 4 1 Daily Care And Cleaning Of First Step P...

Страница 6: ...Table of Contents ii First Step Plus Operations Manual ...

Страница 7: ...atient safety WARNING Serious injury or death can result from the use potential entrapment or non use potential patient falls of Side Rails or other restraints See related Safety Tips section Patient Migration Specialty surfaces have different shear and support characteristics than conventional surfaces and may increase the risk of patient movement sinking and or migration into hazardous positions...

Страница 8: ...conditions regularly especially at bony prominences and areas where moisture or incontinence may occur or collect and consider adjunct or alternative therapies for high acuity patients Early intervention may be essential to preventing serious skin breakdown Fluids Avoid spilling fluids on Air Supply Unit If spills do occur clean fluids from bed and Air Supply Unit wearing rubber gloves to avoid an...

Страница 9: ... mattress with hose connections and tote bag at foot of mattress with elastic harness on underside of overlay Slide elastic harness over each end of mattress Tuck tote bag under foot end of mattress Wrap elastic halter straps located at each end of overlay under and around end of bed mattress 3 Install First Step Plus GORE TEX Cover Sheet Draw sides of sheet down each side of bed and tuck under ma...

Страница 10: ...irflow control knobs located on front of air supply unit to a low setting 10 Transfer patient as described below PATIENT TRANSFER TO FIRST STEP PLUS PRODUCT Ambulatory Patient Transfer 1 Adjust height of both surfaces to comfortable levels for patient ingress and egress 2 Lower Side Rails when applicable 3 Lock caster wheel brakes and caster wheel steering on both surfaces 4 Assist patient in ambu...

Страница 11: ...for individual patient See Air Pressure Adjustment section below 9 Raise and lock Side Rails on both sides of bed If decision has been made to use Side Rails Refer to Risks and Precautions and Safety Tips section in the Introduction chapter of this manual 10 With air pressures set verify patient comfort 11 Lower bed height to lowest level 12 Activate warmer and adjust temperature if desired AIR PR...

Страница 12: ...t section COMPLETION OF PATIENT PLACEMENT 1 Adjust head of bed for patient comfort 2 With air pressures set verify patient comfort 3 Lower bed height to lowest level 4 Activate Warmer and adjust temperature if desired 5 Verify caster wheel brakes and caster wheel steering is locked 6 Assess entrapment and fall hazards and carefully consider the use of bolsters positioning aids low beds floor pads ...

Страница 13: ...es if used refer to Safety Tips to their original position PATIENT TRANSFER TO FIRST STEP PLUS PRODUCT Refer to Patient Placement chapter of this manual if patient is being transferred to First Step Plus product for the first time PATIENT TRANSFER FROM FIRST STEP PLUS PRODUCT Ambulatory Patient Transfer 1 Press green Power Button to turn Off Air Supply Unit Surface will begin to deflate 2 Adjust h...

Страница 14: ...s overlay to maximum by turning all three air flow control knobs to a high setting to provide a firm surface for patient bathing 4 Lower Side Rail on caregivers side CAUTION To reduce risk of patient falls during bathing caregiver must never leave patient unattended or turn away during bathing 5 Gently log roll patient toward opposite Side Rail 6 Bathe and rinse patient s exposed side and posterio...

Страница 15: ...ition 8 Raise and lock Side Rails If decision has been made to use Side Rails Refer to Risks and Precautions and Safety Tips section in the Introduction chapter of this manual 9 Adjust head of bed for patient comfort and return bed and accessories to original configuration BEDPAN REMOVAL 1 Lower head of bed to a level position 2 Lower Side Rail on caregivers side 3 Grasp bedpan firmly with one han...

Страница 16: ...original configuration INCONTINENCE DRAINAGE 1 First Step Plus Gore Tex cover sheet should be placed over mattress overlay 2 Dri Flo breathable underpads are recommended for incontinent patients NOTE Do not use plastic backed underpads Plastic will block moisture vapor transmission and air flow from mattress overlay Dri Flo breathable underpads are absorbent and do not block air flow 3 Watch for i...

Страница 17: ...up and locked 2 Inflate surface fully by adjusting all three air flow control knobs to their highest setting 3 If Side Rails are used lower Side Rail only on Caregiver s side of bed 4 Log roll patient to opposite side of surface 5 Clean patient as required NOTE Avoid spilling fluids on any part of the air supply unit Fluids remaining on the electronic controls can cause corrosion which may cause t...

Страница 18: ...First Step Plus Operations Manual 4 2 Care and Cleaning ...

Страница 19: ...control knob The knobs are labeled and color coded 2 Adjust each surface control knob until optimum pressures and patient comfort has been obtained 3 Adjust air flow so patient is cradled in a slight depression NOTE The Air Supply Unit has a thermal cutoff that will automatically deactivated the unit after a period of time if the air intake is blocked or if the unit is operated with all 3 airflow ...

Страница 20: ...First Step Plus Operations Manual 5 2 Operating Instructions ...

Страница 21: ...rlay inflated 82 L x 35 W x 5 H Air Supply Unit 12 W x 10 3 4 H x 7 D Weight Mattress Overlay inflated 3 lbs Air Supply Unit 11 lbs Electrical Voltage 115 VAC Frequency 60Hz Power 1 5 Amps Max Electrical Leakage 50 Micro Amps Power Cord Length 16 Feet Hospital Grade Specifications subject to change without notice Rev B 05 05 ...

Страница 22: ...First Step Plus Operations Manual 6 2 Specifications ...

Страница 23: ...A Inc P O Box 659508 San Antonio Texas 78265 9508 Call Toll Free 1 800 275 4524 FAX 1 800 275 3417 http www kci1 com GORE TEX is a registered trademark of W L Gore and Associates All other trademarks and service marks designated herein are property of KCI Licensing Inc and it s affiliates This product is patented and or subject to pending patent 2005 KCI Licensing Inc All Rights Reserved 05 05 P N...

Страница 24: ...First Step Plus Operations Manual 7 2 Questions and Information ...