First Step

®

 Plus™ Operations Manual

2-2

Patient Placement

®

5.

Attach hoses to Air Supply Unit:

Locate Hose Set Coupler that holds three hose ends together.

Turn hose set coupler such that hoses are stacked vertically and the blue 
HEAD hose is on top.

 

Insert top tab of coupler into horizontal slot on the right side of the air 

supply unit. Pivot coupler downward, placing hose ends over air outlets 
until bottom of coupler clicks into locked position.

NOTE:

Ensure hose set is not twisted.

6.

Attach hoses to Air Cushion. 

Match color-coded hose ends to their respective patches on air cushions.

Attach hoses snugly by slightly turning while pushing hose end onto overlay hose connectors.

7.

Plug Air Supply Unit into a 115 VAC wall outlet.

NOTE:

Be sure power to this outlet is not controlled by a wall switch.

8.

Activate Air Supply Unit by depressing green Power Switch.

9.

Adjust Head, Body and Foot airflow control knobs, located on front of air supply unit to a low setting.

10.

Transfer patient as described below:

PATIENT TRANSFER TO FIRST STEP

®

 PLUS™ PRODUCT

Ambulatory Patient Transfer

1.

Adjust height of both surfaces to comfortable levels for patient ingress and egress.

2.

Lower Side Rails (when applicable). 

3.

Lock caster wheel brakes and caster wheel steering on both surfaces.

4.

Assist patient in ambulatory transfer, following all applicable safety rules and institution protocols.

5.

Center patient side-to-side on surface. Patients sacrum should be centered head-foot on seat section 
cushion.

6.

Raise and lock Side Rails on both sides of bed. If decision has been made to use Side Rails (Refer to 

Risks 

and Precautions

 and 

Safety Tips

 section in the 

Introduction

 chapter of this manual).

7.

Adjust air flow control knobs to the proper inflation for individual patient. See Air Pressure Adjustment section 
below.

8.

Check for good placement, slide hand flat between bottom of underlying surface and patients buttocks. There 
should be 1" to 1 1/2" between patient and underlying surface.

Rev. B 05/05

Содержание FIRST STEP Plus

Страница 1: ...KCI USA Inc San Antonio Texas 78265 9508 Call Toll Free 1 800 275 4524 Fax www kci1 com P O Box 659508 1 800 275 3417 R TM Plus ...

Страница 2: ......

Страница 3: ... out only by qualified personnel authorized by KCI The electrical installation of the room must comply with the appropriate electrical wiring standards The product must be used in accordance with this manual this product s Maintenance Manual and Quick Reference Guide and all applicable labeling The only servicing to be carried out by the user is described in the Maintenance Chapter of the First St...

Страница 4: ...First Step Plus Operations Manual ...

Страница 5: ...irst Step Plus Product 3 1 Patient Transfer From First Step Plus Product 3 1 Ambulatory Patient Transfer 3 1 Non Ambulatory Patient Transfer 3 1 Patient Transport 3 2 Patient Bathing 3 2 Bedpan Placement 3 3 Bedpan Removal 3 3 Repositioning Patient Using Dri flo Underpads 3 5 Incontinence Drainage 3 5 Skin Care 3 5 General Operation 3 5 CARE AND CLEANING 4 1 Daily Care And Cleaning Of First Step P...

Страница 6: ...Table of Contents ii First Step Plus Operations Manual ...

Страница 7: ...atient safety WARNING Serious injury or death can result from the use potential entrapment or non use potential patient falls of Side Rails or other restraints See related Safety Tips section Patient Migration Specialty surfaces have different shear and support characteristics than conventional surfaces and may increase the risk of patient movement sinking and or migration into hazardous positions...

Страница 8: ...conditions regularly especially at bony prominences and areas where moisture or incontinence may occur or collect and consider adjunct or alternative therapies for high acuity patients Early intervention may be essential to preventing serious skin breakdown Fluids Avoid spilling fluids on Air Supply Unit If spills do occur clean fluids from bed and Air Supply Unit wearing rubber gloves to avoid an...

Страница 9: ... mattress with hose connections and tote bag at foot of mattress with elastic harness on underside of overlay Slide elastic harness over each end of mattress Tuck tote bag under foot end of mattress Wrap elastic halter straps located at each end of overlay under and around end of bed mattress 3 Install First Step Plus GORE TEX Cover Sheet Draw sides of sheet down each side of bed and tuck under ma...

Страница 10: ...irflow control knobs located on front of air supply unit to a low setting 10 Transfer patient as described below PATIENT TRANSFER TO FIRST STEP PLUS PRODUCT Ambulatory Patient Transfer 1 Adjust height of both surfaces to comfortable levels for patient ingress and egress 2 Lower Side Rails when applicable 3 Lock caster wheel brakes and caster wheel steering on both surfaces 4 Assist patient in ambu...

Страница 11: ...for individual patient See Air Pressure Adjustment section below 9 Raise and lock Side Rails on both sides of bed If decision has been made to use Side Rails Refer to Risks and Precautions and Safety Tips section in the Introduction chapter of this manual 10 With air pressures set verify patient comfort 11 Lower bed height to lowest level 12 Activate warmer and adjust temperature if desired AIR PR...

Страница 12: ...t section COMPLETION OF PATIENT PLACEMENT 1 Adjust head of bed for patient comfort 2 With air pressures set verify patient comfort 3 Lower bed height to lowest level 4 Activate Warmer and adjust temperature if desired 5 Verify caster wheel brakes and caster wheel steering is locked 6 Assess entrapment and fall hazards and carefully consider the use of bolsters positioning aids low beds floor pads ...

Страница 13: ...es if used refer to Safety Tips to their original position PATIENT TRANSFER TO FIRST STEP PLUS PRODUCT Refer to Patient Placement chapter of this manual if patient is being transferred to First Step Plus product for the first time PATIENT TRANSFER FROM FIRST STEP PLUS PRODUCT Ambulatory Patient Transfer 1 Press green Power Button to turn Off Air Supply Unit Surface will begin to deflate 2 Adjust h...

Страница 14: ...s overlay to maximum by turning all three air flow control knobs to a high setting to provide a firm surface for patient bathing 4 Lower Side Rail on caregivers side CAUTION To reduce risk of patient falls during bathing caregiver must never leave patient unattended or turn away during bathing 5 Gently log roll patient toward opposite Side Rail 6 Bathe and rinse patient s exposed side and posterio...

Страница 15: ...ition 8 Raise and lock Side Rails If decision has been made to use Side Rails Refer to Risks and Precautions and Safety Tips section in the Introduction chapter of this manual 9 Adjust head of bed for patient comfort and return bed and accessories to original configuration BEDPAN REMOVAL 1 Lower head of bed to a level position 2 Lower Side Rail on caregivers side 3 Grasp bedpan firmly with one han...

Страница 16: ...original configuration INCONTINENCE DRAINAGE 1 First Step Plus Gore Tex cover sheet should be placed over mattress overlay 2 Dri Flo breathable underpads are recommended for incontinent patients NOTE Do not use plastic backed underpads Plastic will block moisture vapor transmission and air flow from mattress overlay Dri Flo breathable underpads are absorbent and do not block air flow 3 Watch for i...

Страница 17: ...up and locked 2 Inflate surface fully by adjusting all three air flow control knobs to their highest setting 3 If Side Rails are used lower Side Rail only on Caregiver s side of bed 4 Log roll patient to opposite side of surface 5 Clean patient as required NOTE Avoid spilling fluids on any part of the air supply unit Fluids remaining on the electronic controls can cause corrosion which may cause t...

Страница 18: ...First Step Plus Operations Manual 4 2 Care and Cleaning ...

Страница 19: ...control knob The knobs are labeled and color coded 2 Adjust each surface control knob until optimum pressures and patient comfort has been obtained 3 Adjust air flow so patient is cradled in a slight depression NOTE The Air Supply Unit has a thermal cutoff that will automatically deactivated the unit after a period of time if the air intake is blocked or if the unit is operated with all 3 airflow ...

Страница 20: ...First Step Plus Operations Manual 5 2 Operating Instructions ...

Страница 21: ...rlay inflated 82 L x 35 W x 5 H Air Supply Unit 12 W x 10 3 4 H x 7 D Weight Mattress Overlay inflated 3 lbs Air Supply Unit 11 lbs Electrical Voltage 115 VAC Frequency 60Hz Power 1 5 Amps Max Electrical Leakage 50 Micro Amps Power Cord Length 16 Feet Hospital Grade Specifications subject to change without notice Rev B 05 05 ...

Страница 22: ...First Step Plus Operations Manual 6 2 Specifications ...

Страница 23: ...A Inc P O Box 659508 San Antonio Texas 78265 9508 Call Toll Free 1 800 275 4524 FAX 1 800 275 3417 http www kci1 com GORE TEX is a registered trademark of W L Gore and Associates All other trademarks and service marks designated herein are property of KCI Licensing Inc and it s affiliates This product is patented and or subject to pending patent 2005 KCI Licensing Inc All Rights Reserved 05 05 P N...

Страница 24: ...First Step Plus Operations Manual 7 2 Questions and Information ...

Отзывы: