Chapter 6. Calibrator Operation
30
Kaye LTR -90 Operating Manual
6.1.2b
Set the Temperature with a Preset SETPOINT
The Calibrator stores eight user-configured preset SETPOINT temperatures to help quickly set a temperature. To
change a SETPOINT preset see “Change a Preset SETPOINT” below.
To load and use a preset SETPOINT (see Figure 15 below):
1.
Push
on the Main screen.
2.
Push .
3.
Push
or
to move between presets (1 through 8).
4.
Push
to select a preset.
Figure 15: Preset Selection
6.1.2c
Change a Preset SETPOINT
To change a preset SETPOINT (see Figure 15 above):
1.
Push
on the Main screen.
2.
Push .
3.
Push .
4.
Push
or
to move between presets (1 through 8). Highlight the preset to change, then push
.
5.
Push
and
to highlight a digit, then push
to increase the digit or push
to decrease the digit.
6.
Push
to save the preset.
6.1.3 Cancel or Stop Temperature Change
To cancel or stop temperature change:
1.
Push
on the Main screen.
2.
Push
.
66.03°C
SELECT PRESET
EDIT
PRESET
PRESET:
(SETPOINT: 80.00
°C
)
Содержание LTR -90
Страница 4: ...ii no content intended for this page...
Страница 14: ...Chapter 2 Specifications 8 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 24: ...Chapter 4 Parts and Controls 18 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 27: ...Kaye LTR 90 Operating Manual 21 Chapter 5 General Operation Figure 7 Insert Installation and Removal...
Страница 29: ...Kaye LTR 90 Operating Manual 23 Chapter 5 General Operation Figure 8 Probe Insertion Figure 9 Rubber Insulator...
Страница 30: ...Chapter 5 General Operation 24 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 44: ...Chapter 6 Calibrator Operation 38 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 49: ......