Kaye LTR -90 Operating Manual
7
Chapter 2. Specifications
IRTD hole: 6.7 mm (0.265 in)
Sensor hole: 9 mm (0.354 in)
Heating Time
[1]
-95°C to 140°C: 40 min
Cooling Time
[1]
23°C to -90°C: 80 min
23°C to -95°C: 90 min
140°C to 23°C: 60 min
Stabilization Time
[2]
15 min
Resolution
0.01°
Display
LCD, °C or °F user selectable
Size (H x W x D)
480 mm x 205 mm x 380 mm (18.8 in x 8.0 in x 14.9 in)
Weight
16 kg (35 lb)
Power Requirements
100 V to 115 V (±10%) 50/60 Hz, 575 W (X0395, X0395JP)
200 V to 230 V (±10%) 50/60 Hz, 575 W (X0395CE, X0395AS)
System Fuse Ratings
115 V: 6.3 A T 250 V (Kaye part #:193-100)
230 V: 3.15 A T 250 V (Kaye part#: 193-099)
Safety
IEC 61010-1, Installation Category II, Pollution degree 2
Electromagnetic Environment
IEC 61326-1: Basic
Refrigerants
R32 (Difluoromethane)
< 20 g, ASHRAE Safety Group A2L
R704 (Helium)
< 20 g, ASHRAE Safety Group A1
Notes
:
[1] — For ambient temperature of 23°C.
[2] —Time from when the SETPOINT is reached to when the unit is within Stability specification.
Содержание LTR -90
Страница 4: ...ii no content intended for this page...
Страница 14: ...Chapter 2 Specifications 8 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 24: ...Chapter 4 Parts and Controls 18 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 27: ...Kaye LTR 90 Operating Manual 21 Chapter 5 General Operation Figure 7 Insert Installation and Removal...
Страница 29: ...Kaye LTR 90 Operating Manual 23 Chapter 5 General Operation Figure 8 Probe Insertion Figure 9 Rubber Insulator...
Страница 30: ...Chapter 5 General Operation 24 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 44: ...Chapter 6 Calibrator Operation 38 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 49: ......