| Kaye LTR-150 - Dry Block & Liquid Bath All-in-One Temperature Calibrator
Set Cycles:
You can set the number of cycles here. For every cycle, the test points in the selection list are
approached successively.
Tap on the parameter field to change the number of cycles.
Input the desired number in the input window and confirm the input.
*
The new values are adopted and you will be returned to the previous window.
Set Steps:
Tap on the parameter field Number of Steps to change the number of steps.
Input the desired number and confirm the input.
*
The number of steps is adopted and the interval is calculated afresh.
Set Dwell Time:
Tap on the parameter field Dwell Time, input the new values and confirm the input.
*
The new value is adopted and you will be returned to the previous window.
One-Sided Step Selection:
Activate the parameter field by tapping if the test points are to be approached in succession
from the first to the last test point.
Two-Sided Step Selection:
Activate this parameter field by tapping if the test points are to be approached in succession
from the first to the last back to the first test point.
Select Gradient:
Tap on the parameter field Gradient, input the new values and confirm the input.
*
The new value is adopted and you will be returned to the previous window.
At the completion of configuration of the test task, continue to tap the check mark icon
thru all screens until you return to the select test task screen.
To load, select desired test task and tap the check mark icon. The test task will load
and screen will return to Standalone main screen with name of selected test task.
Chapter 7. Standalone Manual Mode