background image

www.kawemed.com

9

K5 is the point at which the sounds heard with a stethoscope are 
no longer audible.
K4 is the point at which the sounds heard with a stethoscope 
vary from clear tapping to muffled.

Measuring blood pressure: 

 Attaching the arm cuff (instructions given for right-
handed users):

1.   After pushing back the shirt sleeve or otherwise freeing 

the upper arm of clothing, open the cuff‘s Velcro strap and 
pull the cuff up over the upper left arm. The tube should 
be positioned so that to point downward. The lower edge 
of the cuff should stay approx. 2-3 cm above the elbow.

2.   The artery pointer on the cuff (KaWe logo must be visible 

on the outside) must be placed on the inside of the 
upper arm near the palpating artery. If this artery cannot 
be seen, it can be found by feeling the pulse with the 
fingertips. 

3.   For the doctor’s blood pressure measuring device, a 

stethoscope is required in addition. The stethoscope chest 
piece (membrane side) must be inserted under the cuff 
and positioned over the artery. 

4.  Make sure that the cuff is always at the heart level. 
5.   Using the Velcro closure, fasten the cuff firmly but not 

tightly (leaving about 1-2 cm of space). 

6.   Prepare the stethoscope for use (insert the ear pieces into 

your ears).

Creating pressure: 

1.   Close the air-release valve on the hand pump by turning 

the screw to the right. Do not over-tighten the screw. 

2.  Inflate the cuff quickly and rhythmically with the hand 

pump and use the diaphragm on the stethoscope to listen 
to the pulse. 

3.   During this process, watch the device gauge and hold the 

left arm absolutely still. 

4.   If a pulse can no longer be heard with the stethoscope, 

raise the pressure in the cuff once again with the hand 
pump by 30 mmHg.

Releasing pressure:  

1.   Open the air release valve slowly, so that the drop in 

pressure on the gauge is approx. 2-3 mmHg/s.

Note: The speed of the pressure release affects the exactness 
of measurement results. The pressure must not be released too 
quickly or the point at which the pulse restarts will be missed. 
If the pressure is released too slowly, the arm will be left too 
long without the supply of blood because the pressure cuts 
off the blood flow.

Determining the systolic pressure: 

1.   While releasing the pressure, listen continuously for the 

sound of the pulse in the artery with the stethoscope. 
When the first pulse is heard, read the upper blood 
pressure value (systolic pressure) and remember this value.

 2.   As the pressure is released even further, the pulses 

will become quieter until they disappear completely. 
Remember the last pressure value by which a pulse was 
heard. This is the lower blood pressure value (diastolic 
pressure).

 

Completing the measurement procedure: 

 1.   Open the air release valve completely to let the air escape 

completely from the cuff. 

2.   Remove the cuff from the arm and store the entire 

instrument in its storage case.

Cleaning:

Gauge: 

To clean the device, use only a soft, dry cloth.

Reusable cuff: 

If necessary, wipe the cuff with a damp cloth. 

Disinfection: 

Gauge: 

For disinfection of the device by wiping (minimum 

exposure time of at least 5 minutes) it is recommended to use 
the disinfectant Mikrozid sensitive liquid (manufactured by 
Schülke & Mayr). 

ENGLISH

Отзывы: