background image

www.kawemed.com

8

Dear Customer, thank you for choosing our innovative aneroid 
blood pressure monitor. Our products are known for their 
high quality and longevity. This KaWe product meets the 
requirements of EС directive 93/42/EEC (standards for medical 
products) and DIN EN ISO 81060-1:2012 standard “Non-
invasive sphygmomanometers - Part 1: Requirements and test 
methods for non-automated measurement type”.

Please read these instructions thoroughly and 

completely before use and follow the given care 
instructions.

Use: 

This device and its accessories are only to be used by 
physicians and trained professionals for monitoring blood 
pressure.

Purpose: 

This device is intended for the non-invasive measurement of 
systolic and diastolic blood pressure.

Unsuited use/contraindication of 

the device: 

Any use other than that described here shall be deemed 
improper. The manufacturer shall not be liable for any resulting 
damages. The user alone shall bear all risks.

Features/preparation:  

This blood pressure measuring device is an aneroid measuring 
device. During measurement, there is a build-up of pressure 
that constricts an artery in the upper arm. As the pressure is 
released, the upper blood pressure reading (systolic blood 
pressure) is calculated and after a complete blood flow 
through the artery, further pressure is released and thereby 
the lower blood pressure reading (diastolic blood pressure) 

User

s Manual

Aneroid blood pressure measuring device (for doctors)

is calculated.

1.   Familiarize yourself fully with this product before using it. 
2.   Do not take your blood pressure directly after bodily 

exertion or in a state of excitement.

 3.   Relax for approx.15 min in order to avoid measurement 

inaccuracies.

 4.   The measuring time shall not exceed two minutes. 

Between two successive measurements, a break must be 
taken for a minimum of two minutes. 

5.   In order to ensure correct measuring results, make sure 

that the proper cuff size is used. Different cuff sizes 
are available for different applications. Please consult 
an authorized dealer. The single-tube cuff only fits 
MASTERMED A1 and A1+. The dual-tube cuff is used with 
MASTERMED A2, A2+ and A3.

Basic rules for patients when 

measuring blood pressure:

-   sit comfortably
-   do not cross your legs 
-   lean back or find other support for your back and arms
-   the middle of the cuff should be on the upper arm at the 

heart level

-   during blood pressure measurement, relax and do not talk 

if possible

-   rest for 5 minutes before the first use

Recommended for auscultation method:
-   in adults, phase V (K5)
-   in children aged 3 to 12 years, phase IV (K4)
-   in pregnant women, phase V (K5), unless noises are heard 

in the cuff when pressure reduces, in such case phase IV 
(K4) of Korotkoff sounds is used.

ENGLISH