9
Details about electromagnetic compatibility according to
EN60601-1-2
9.1
Electromagnetic Transmissions
The Status 1080 TM treatment unit is for use in an environment like the one cited
below. The customer or user of the Status 1080 TM should ensure that it is used in
the correct environment.
Measurements of noise transmis‐
sions
Conformance
Electromagnetic environment -
Guidelines
HF transmission according to
CISPR 11
Group 1
The Status 1080 TM uses HF ener‐
gy only for its internal operation. Its
HF transmission is therefore very
low, and it is improbable that neigh‐
bouring electronic devices will be
disturbed.
HF transmission according to
CISPR 11
Class B
The Status 1080 TM is for use in all
facilities including residential ones,
and facilities that are directly con‐
nected to a public power supply that
also supplies residential buildings.
Transmissions of harmonics ac‐
cording to IEC 61000-3-2
Class A
The Status 1080 TM is for use in all
facilities including residential ones,
and facilities that are directly con‐
nected to a public power supply that
also supplies residential buildings.
Transmission of voltage fluctuations
or flicker according to IEC
61000-3-3
Conforms
The Status 1080 TM is for use in all
facilities including residential ones,
and facilities that are directly con‐
nected to a public power supply that
also supplies residential buildings.
9.2
Resistance to electromagnetic interference
The Status 1080 TM treatment unit is for use in an environment like the one cited
below. The customer or user of the Status 1080 TM should ensure that it is used in
the correct environment.
Immunity tests
IEC 60601 test level
Conformance level
Electromagnetic environ‐
ment - guidelines
Electrostatic discharge
(ESD) according to IEC
61000-4-2
± 6 kV contact discharge
± 8 kV atmospheric dis‐
charge
± 2/4/6 kV contact dis‐
charge
± 2/4/8 kV atmospheric
discharge
Floors should be made of
wood or concrete or have
ceramic tiles. When the
floor is made of synthetic
material, the relative hu‐
midity must be at least
30%.
Fast transient electrical
disturbances/ Bursts ac‐
cording to IEC 61000-4-4
± 2 kV for power lines
± 1 kV for input and output
lines
± 2 kV for power lines
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment.
Instructions for use Status 1080 TM
9 Details about electromagnetic compatibility according to EN60601-1-2 | 9.1 Electromagnetic Transmissions
72/74