The chair moves into the automatic position.
You can also use the cross switch.
Lock the chair functions when the instrument is removed
▶ Press the foot pedal.
The chair functions are locked when an instrument is removed.
4.3.6
Safety shut-off
To prevent collisions arising from the movement of the patient chair, safety shutoff
switches are installed to protect the patient and practice personnel from injury and
the treatment unit from damage.
CAUTION
Despite the safety shutoff, collisions with the patient chair can occur when the
assistant's unit is in certain positions.
Damage to the assistant element and patient chair.
▶ Keep the assistant unit out of the range of movement of the patient chair.
▶ Always monitor the chair movement.
CAUTION
Pinching from the treatment chair.
The safety shutoff of the treatment chair is activated by lifting the respective com‐
ponent. Depending on the patient's body weight and the leverage, more force can
be exerted on the object to be triggered than is necessary to trigger the switching
function.
▶ The treatment personnel must move outside of the chair's swinging range
whenever the chair moves.
Instructions for use Status 1080 TM
4 Operation | 4.3 Adjusting the dental chair
42/74