Instructions for use SONICflex seal tips
2 Safety | 2.1 Infection hazard
7 / 20
2 Safety
The instructions for use are a component of the product and must be read care-
fully prior to use and be accessible at all times.
The device may only be used in accordance with the intended use, any other
type of use is not permitted.
2.1 Infection hazard
Patients, users or third parties could be infected by contaminated medical
devices.
4
Take suitable personal protective measures.
4
Follow the instructions for use of the components.
4
Before initial startup and after each use, reprocess the product and ac-
cessories appropriately.
4
Carry out the reprocessing as described in the instructions for use. The pro-
cedure has been validated by the manufacturer.
4
If you deviate from this validated procedure, make sure that the sterilisation
procedure is effective.
4
Reprocess the product and accessories appropriately before disposal.
4
In the case of injury to soft tissue, do not continue treatment in the oral
cavity with instruments driven by compressed air.
4
Place the torque wrench on the tip when the handpiece is in its holder.
2.2 Technical condition
A damaged device or components could injure patients, users and third parties.
4
Only operate devices or components if they show no signs of damage on the
outside.
4
Check that the device is working properly and is in satisfactory condition be-
fore each use.
Using the product at improper or excessive power settings can lead to tip
breakage and ensuing injury.
4
Do not select improper or excessive power settings.
4
Check the tip for secure attachment before treatment.
Fracture may occur as a result of long-term use or damage (dropped to the
floor or mechanical change of the original shape).
4
Before each use, check the operational safety of the tips by gently pressing
on them with your thumb or forefinger.
4
In addition, expose the tips to approx. 10 N (1 kg) mechanical load without
function.
Worn tips can fracture or be contaminated and thus cause injury or infection.
4
SONICflex tips need to be replaced every 9 to 12 months.
4
Have parts with sites of breakage or surface changes checked by the service
personnel.
4
Safety checks may only be performed by trained service personnel.
4
If the following defects occur, stop working and have the service personnel
carry out repair work:
▪ Malfunctions
▪ Damage
▪ Irregular running noise
▪ Excessive vibration
▪ Overheating
▪ Tip is not seated firmly in the handpiece