3. Using the system
3.3 Preparing an IP for exposure
IMPORTANT NOTICE:
If the IP is being used for the very
fi
rst time
or if it has not been used within the last 24
hours, it must be erased before use to re-
move any fogging caused by background ra-
diation. See section,
3.11 IP erasing mode
(Initial erasing of the IPs)
.
CAUTION:
When handling IPs, protective covers and hy-
giene bags take all appropriate measures and
precautions to prevent cross contamination.
NOTICE:
If you are taking an occlusal projection expo-
sure refer to section
3.12 Taking an occlu-
sal projection image
and to the separate
instructions supplied with
the
Comfort Occlusal™ 4C
kit.
1. Place the IP you wish to use onto a pro-
tective cover. The
light blue
side (sensi-
tive) of the IP must face and be placed on
the half of the protective cover that is the
same shape as the IP.
2. Fold the half of the protective cover with
the semicircular cutout over onto the IP.
The metal disk on the back (black side) of
the IP must appear in the semicircular cut
out.
3. Turn the protective cover and IP over so
that the black side of the protective cover
is uppermost. This makes it easier to slide
the protective cover and IP into the hy-
giene bag.
Содержание SCAN EXAM
Страница 1: ...SCAN EXAM Digital imaging plate scanner User Manual ENGLISH 215948 rev 1 0 805 5059...
Страница 2: ......
Страница 4: ...IV User manual 215948...
Страница 7: ...User manual 215948 VII...
Страница 8: ...VIII User manual 215948...
Страница 12: ...4 SCAN EXAM 2 Intraoral imaging plate unit 2 Intraoral imaging plate unit 2 1 Main parts and controls...
Страница 49: ...SCAN EXAM A 3 A Technical Specifications A 3 Main dimensions...
Страница 54: ...A 8 SCAN EXAM A Technical Specifications...