7 Maintenance
7.
The calibration image appears on the GUI.
8.
Acknowledge the calibration result by pressing the
OK
button.
7.3
Quality control
The quality control (QC) programs in the Quality control menu are used to ensure that the image quality
remains constant.
Quality control should be performed at regular intervals, preferably at least once a month and always
after calibration.
NOTICE!
The device will automatically remind of retaking the QC images. If you want the change
the frequency of the reminders, contact your service for assistance.
NOTICE!
QC programs produce X-rays. Protect yourself from radiation.
7.3.1
PAN QC
NOTICE!
The PAN QC is an optional, yet recommendable, procedure and is mandatory to be
performed where local regulations require it.
1.
GUI:
Go to device settings.
2. GUI:
Select
PAN QC
program from the
Quality control
menu.
ORTHOPANTOMOGRAPH
™