Function test
▶ Insert a tool (cutter or grinder) and turn it radially by hand in various directions,
applying slight thumb pressure, to precisely align the O-rings.
▶ Start the medical device.
▶ Do a test run under axial and radial load with no patient contact.
▶ Tighten the cover again with a torque of 100 Ncm ± 6 Ncm.
Note
If the unit runs poorly or heats up in untypical manner, please return the entire
turbine contra-angle handpiece to KaVo for testing.
Instructions for use Spare cartridge for MASTERtorque - 2.000.2288, 2.000.2266