▶ In the detector adjustment menu, press the START key to start adjustment.
▶ The user will be guided through the detector adjustment by the program. Follow
the instructions on the touchscreen.
4.16.3
Determining the zero value
The following instruction appears on the touchscreen: "Point the handpiece to the
floor".
Note
It should be ensured that the handpiece is not directed at a foreign light source.
After 5 seconds, the measurement of the zero value begins automatically.
Duration of the measurement is 2 seconds.
4.16.4
Determine reference value
The following instruction appears on the touchscreen: "Hold instrument on reference
probe".
After 5 seconds, the measurement of the reference value begins automatically.
Duration of the measurement is 2 seconds.
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
4 Operation | 4.16 Detector adjustment
54/76
Содержание KEY Laser 3+ 1343
Страница 1: ...Instructions for use KEY Laser 3 1343 KEY Laser III 1243 upgraded Always be on the safe side ...
Страница 79: ......
Страница 80: ...1 007 0175 kf 20100225 03 en ...