4.3 Checking the pressures
CAUTION
Compressed air connection on devices.
Contaminated and humid compressed air leads to premature wear and
tear.
▶ Ensure that the cooling air is dry, clean and uncontaminated in accord‐
ance with EN ISO 7494-2.
A minimum drive pressure of 2.5 bar (29 psi) is required to operate the
medical device.
3.0 bar (44 psi) is ideal.
The air consumption is 50 to 60 Nl/min.
4 Start up and shut down