7 Preparation methods according to ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from non-sterile products.
There is a risk of infection from contaminated medical devices.
▶ Observe suitable personal protective measures.
▶
Remove all residual cement, composite or blood without delay.
▶
The medical device must be dry when transporting it to the site of re‐
processing.
▶
Do not place it in a solution or the like.
▶
Reprocess the medical device as soon as possible after treatment.
Preparation methods according to ISO 17664
35