(Depending on the available autoclave,) select a suitable procedure from
the following sterilisation processes:
▪ Autoclave with three times initial vacuum:
–
at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4
°F)
▪ Autoclave using the gravitation method:
–
at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4
°F) or alternatively
–
at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4
°F)
▶
Use according to the manufacturer's Instructions for Use.
7.8 Storage
Prepared products must be stored, protected from germs (as far as possi‐
ble) and dust, in a dry, dark, cool room.
Setup methods according to DIN EN ISO 17664
55