Depending on the device, select a suitable process from the following ster‐
ilisation processes.
Autoclave with a triple pre-vacuum for least four minutes at 134°C ± 1°C
(273°F ± 1.8°F)
Autoclave with gravitation process for least ten minutes at 134°C ± 1°C
(273°F ± 1.8°F)
Autoclave with gravitation process for at least 60 minutes at 121°C ± 1°C
(250°F ± 1.8°F)
Follow the manufacturer's instructions for use.
6.10 Storage
Prepared products must be stored, protected from germs (as far as possi‐
ble) and dust, in a dry, dark, cool room.
Preparation methods according to ISO 17664