![KaVo INTRA shank C 09 Скачать руководство пользователя страница 33](http://html2.mh-extra.com/html/kavo/intra-shank-c-09/intra-shank-c-09_instructions-for-use-manual_5096370033.webp)
▶
Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three times
for 2 seconds each time. Remove the medical device from the spray
attachment and let the cleaner work for one minute.
▶
Afterwards, rinse for 3-5 seconds with KaVo DRYspray.
See also: KaVo CLEANspray / KaVo DRYspray Instructions for Use
Note
KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are
only available in the following countries:
Germany, Austria, Switzerland, Italy, Spain, Portugal, France, Luxem‐
bourg, Belgium, Netherlands, United Kingdom, Denmark, Sweden, Fin‐
land and Norway.
In other countries, only automated interior cleaning with thermodisinfec‐
tors in accordance with ISO 15883-1.
Preparation methods according to ISO 17664