![KaVo INTRA LUX KL 703 LED Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/kavo/intra-lux-kl-703-led/intra-lux-kl-703-led_instructions-for-use-manual_3862780018.webp)
Instructions for use INTRA LUX Motor KL 703 LED – 1.007.0150
6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions before initial start-up
18 / 28
6 Checking for malfunctions and troubleshooting
6.1 Checking for malfunctions before initial start-up
CAUTION
Danger from hot lamp.
Risk of burn injury.
4
Do not touch lamp after previous operation. Allow the lamp to cool down.
CAUTION
Missing or damaged O-rings.
Malfunction and premature failure.
4
Make sure that all O-rings are on the coupling and are undamaged.
CAUTION
Inappropriate attachment of the handpiece to the coupling.
Damage to the O-rings.
4
Take care when you attach the handpiece to the coupling.
NOTE
Stop working if the O-ring is missing or damaged.
4
If the motor overheats while idling, check the amount of cooling air.
4
If the motor overheats on load, check the handpiece also.
4
When the speed drops / is uneven, check the hose connection.
4
Missing O-ring of the motor coupling needs to be replaced.
4
If the intensity of the KaVo Mini LED lamp is low, increase the cold light in-
tensity on the unit.
4
If the KaVo Mini LED lamp glows red or not at all, turn the KaVo Mini LED
lamp by 180° and insert it.
Also refer to:
6.2.1 Replacing the KaVo Mini LED lamp, Page 19
6.2 Troubleshooting
WARNING
Repair WITHOUT using KaVo original spare parts.
Parts such as the cover can come loose and cause injury. Aspiration, swallowing
of parts and danger of suffocation.
4
Only use spare parts that comply with the specification for repair; original
KaVo spare parts comply with the specification.
NOTE
If a repair is done with NON-KaVo original spare parts, this may constitute a
product modification that leads to the loss of CE conformity. In the event of
damage, the responsibility is with the service company or the operator.
The introduction of a modified product in the market, in which the reasonable
suspicion exists to endanger the safety and health of patients or users, is pro-
hibited by Medical Device Law §4, Abs.1 No. 1 and therefore requires its own
conformity check.