Instructions for use INTRA LUX Motor KL 703 LED – 1.007.0150

5 Operation | 5.1 Connecting the motor to the supply hose

16 / 28

5 Operation

NOTE

At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 minutes (without transmission handpieces being at-
tached) and if there is a risk of contamination from reflux or back suction, the
system may also need to be rinsed for 20 to 30 seconds after each patient.

5.1 Connecting the motor to the supply hose

4

Apply a small amount of KaVo Spray to the O-rings on the supply hose.

4

Connect the motor to the supply hose and twist.

ð

The correct attachment position is attained automatically.

4

Tighten the hose-side union nut proceeding in the direction of the arrow.

NOTE

Details on the hose coupling can be obtained from the manufacturer.

5.2 Regulating the spray water

 CAUTION

Hazard from insufficient amount of spray water.

Damage to the cog due to overheating.

4

Make sure that the motor speed is appropriate for the preparation on hand.

4

Use the requisite minimal amount of spray water.

4

Rotate the regulating sleeve of the motor hose to adjust the water portion
of the spray in a continuous manner.

ð

When the regulating sleeve is locked into place, the passage is opened
the maximum amount.

5.3 Attaching an instrument to the motor

All straight and contra-angle handpieces with INTRAmatic connection according
to DIN EN ISO 3964 can be attached.

 CAUTION

Make sure that the handpiece is firmly seated on the motor.

The handpiece detaching inadvertently from the motor during treatment is a
hazard to the patient and user.

4

Pull on the instrument before each treatment to check that it is engaged
and locked in place.

CAUTION

Attaching and removing instruments during rotation.

Property damage to the product.

4

Do not attach or remove the instruments during rotation.

4

Only use reliable handpieces.

4

Place the instrument on the motor, lightly press it against the motor while
turning it in the direction of the arrow until the guide stud can be heard to
lock into place.

4

Pull on it to make sure that the instrument is securely attached to the mo-
tor.

Содержание INTRA LUX KL 703 LED

Страница 1: ...Instructions for use INTRA LUX Motor KL 703 LED 1 007 0150...

Страница 2: ...Distribution KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Страница 3: ...motor to the supply hose 16 5 2 Regulating the spray water 16 5 3 Attaching an instrument to the motor 16 5 4 Removing an instrument from the motor 17 6 Checking for malfunctions and troubleshooting 1...

Страница 4: ...ice If you have any questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service einrichtungen kavo com or service treatmentunits kavo com Please refer to the serial num...

Страница 5: ...es Transportation and storage conditions temperature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from...

Страница 6: ...or use INTRA LUX Motor KL 703 LED 1 007 0150 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if not prevented can lead to m...

Страница 7: ...ore disposal 2 2 Explosion hazard Electrical sparks in the product can lead to explosion or fire 4 Do not use the product in explosion hazard areas 4 Do not operate the product in an oxygen enriched a...

Страница 8: ...of time 2 6 Electromagnetic fields Electromagnetic fields might interfere with the functions of implanted systems such as pacemakers 4 Ask patients if they have a cardiac pacemaker or other system im...

Страница 9: ...als where the medical device is brought to a professional shop for cleaning servicing and a func tion check Define the service interval depending on the frequency of use As a result of the use of NON...

Страница 10: ...fety 2 11 Disposal 10 28 2 11 Disposal This product is subject to the EC directive governing waste electrical and elec tronic devices and must be submitted to special disposal in Europe 4 For further...

Страница 11: ...n areas at risk of explosion Intended only for dental treatment by a dental professional the product must not be modified or used for any other purpose since this may be haz ardous A medical device ac...

Страница 12: ...011 2600 ELECTROmatic TMM TMC Mat no 0 696 0161 KL 2 Mat no 1 004 4317 KL 3 Mat no 1 006 1037 KL 4 Mat no 1 003 3490 ELECTROtorque TLC Mat no 1 001 8748 ELECTROtorque PLUS Motor voltage rating 22 V AC...

Страница 13: ...duct to which the motor is to be attached Check the suitability of the product for operation with the motor NOTE Designed for operation on KaVo units 4 ESTETICA E80 ESTETICA E70 ESTETICA E50 ESTETICA...

Страница 14: ...pressure 700 hPa to 1060 hPa 10 psi to 15 psi 3 3 Transportation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be...

Страница 15: ...to 7 Reprocessing steps in accordance with ISO 17664 Page 20 Currently applicable packaging law Dispose of and recycle the packaging appropriately in accordance with current packaging law employing wa...

Страница 16: ...speed is appropriate for the preparation on hand 4 Use the requisite minimal amount of spray water 4 Rotate the regulating sleeve of the motor hose to adjust the water portion of the spray in a contin...

Страница 17: ...ions for use INTRA LUX Motor KL 703 LED 1 007 0150 5 Operation 5 4 Removing an instrument from the motor 17 28 5 4 Removing an instrument from the motor 4 Pull the instrument off the motor in axial di...

Страница 18: ...n 4 Missing O ring of the motor coupling needs to be replaced 4 If the intensity of the KaVo Mini LED lamp is low increase the cold light in tensity on the unit 4 If the KaVo Mini LED lamp glows red o...

Страница 19: ...k properly Case 1 KaVo Mini LED lamp is on Case 2 The KaVo Mini LED lamp is faint 4 Increase the cold light intensity on the unit until the desired illuminance is reached Case 3 The KaVo Mini LED lamp...

Страница 20: ...films may be formed in the lines This is associated with a risk of infection This results in the following requirements for the workflow at the practice 4 Rinse all withdrawal sites of the water condu...

Страница 21: ...cal device with a second CaviWipes cleaning and disinfection wipe to remove any remaining invisible contaminations Manual external disinfection 4 Finally thoroughly wipe down the outer surfaces of the...

Страница 22: ...rvice regularly with suitable agents CAUTION Excess care products leaking from the contra angle handpiece oil cleanser and disinfectant can penetrate into the motor and cause it to malfunction 4 Pleas...

Страница 23: ...the following sterilisation processes depending on the available steriliser Steriliser with triple pre vacuum at least 3 minutes at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F Autoclave using the gra...

Страница 24: ...or KL 703 LED 1 007 0150 8 Consumables 24 28 8 Consumables Available from specialised dental dealers Material summary Mat No KaVo Mini LED lamp 1 007 8474 O ring 8 3x0 68 0 200 6120 Airflow measuring...

Страница 25: ...or defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or connection instruc...

Страница 26: ......

Страница 27: ......

Страница 28: ...1 008 1048 bd 20220914 06 en...

Отзывы: