Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three times
for 2 seconds each time. Remove the medical device from the spray
attachment and let the cleaner work for one minute.
▶
Afterwards, rinse for 3-5 seconds with KaVo DRYspray.
See also: KaVo CLEANspray / KaVo DRYspray Instructions for Use
Note
KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are
only available in the following countries:
Germany, Austria, Switzerland, Italy, Spain, Portugal, France, Luxem‐
bourg, Belgium, Netherlands, United Kingdom, Denmark, Sweden, Fin‐
land and Norway.
In other countries, only automated interior cleaning with thermodisinfec‐
tors in accordance with ISO 15883-1.
Preparation methods according to ISO 17664
33
Содержание INTRA Chirurgie shank 3624 N
Страница 1: ...Instructions for use INTRA Chirurgie shank 3624 N REF 0 534 5660...
Страница 17: ...3 Product description INTRA surgical shank 3624 N Mat no 0 534 5660 Product description 15...
Страница 50: ...Note Observe the expiration date of the sterilised item Preparation methods according to ISO 17664 48...
Страница 52: ...Material summary Mat no ROTAspray 2142 A 0 411 7520 QUATTROcare plus Spray 2108 P 1 005 4525 Tools 50...
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ...0 589 2112 kb 20100507 05 en...