Autoclave using the gravitational method: sterilize for at least 10 min. at
134°C ± 1°C (273°F ± 1.8°F)
Drying time: 30 min.
Autoclave using the gravitational for at least 60 min. at 121°C ± 1°C (250°F
± 1.8°F)
Drying time: 30 min.
Follow the manufacturer's instructions for use.
6.10 Storage
Prepared products must be stored germ-free and protected from dust in a
dry, dark and cool room.
Note
Observe the expiration date of the sterilized item.
Preparation methods according to ISO 17664
44