5.3 Insert the milling cutters or diamond grinders
Note
Only use carbide cutters or diamond grinders that correspond to ISO
1797-1 type 3, are made of steel or hard metal and meet the following
criteria:
- shaft diameter: 2.334 to 2.350 mm with drill stop
- Overall length: max. 44,5 mm
- Shaft clamping length: min. 31 mm
- Cutting diameter: max. 3 mm
Without drill stop
- Overall length: max. 62.5 mm
- Shaft clamping length: min. 49 mm
- Blade diameter: max. 3 mm