Troubleshooting

5.1

   

Exchanging the O-rings on the coupling on the supply hose

CAUTION

Hazard from improper care of the O-rings.

Malfunctions or complete failure of the medical device.

▶ Do not use Vaseline or other grease or oil.

Note
The O-ring on the coupling may only be lubricated with a cotton ball wet with KA‐

VO spray.

▶ Press the O-ring between your fingers to form a loop.

▶ Shove the O-ring to the front, and remove it.

▶ Insert new O-rings into the grooves.

5.2

   

Cleaning the spray nozzle.

CAUTION

Hazard from insufficient amount of spray water.

Insufficient spray water can cause the medical device to overheat and damage

the tooth.

▶ Check spray water channels and if necessary clean spray nozzles with the

nozzle needle (Mat. no. 0.410.0921).

▶ Clean the water passage in the spray nozzles by using the nozzle needle

(Mat. no. 0.410.0921).

5.3

   

Change the water filter

▶ Unscrew the sleeve ① anticlockwise from the insert ② and pull off.

▶ Pull back the insert ② slightly.

▶ Use tweezers to remove the water filter ③ from the insert ②.

▶ Insert a new water filter.

▶ Pull back the insert ② slightly.

▶ Place the sleeve ① on the insert ②, and screw it tight clockwise.

 

 

 

 

 

 

 

 

 

 

 

Instructions for use GENTLEsilence LUX 8000B - 1.000.0800 | GENTLEsilence 8000C - 1.000.2800
5 Troubleshooting | 5.1 Exchanging the O-rings on the coupling on the supply hose

10/18

Содержание GENTLEsilence LUX 8000B

Страница 1: ...Instructions for use GENTLEsilence LUX 8000B 1 000 0800 GENTLEsilence 8000C 1 000 2800 Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...664 11 6 1 Preparations at the site of use 11 6 2 Cleaning 11 6 2 1 Manual cleaning external 11 6 2 2 Automated external cleaning 11 6 2 3 Manual cleaning internal 12 6 2 4 Automated internal cleaning...

Страница 4: ...r users and service technicians Can be steam sterilised at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Thermodisinfectable CE mark European Community A product bearing this mark meets the requirements of the appl...

Страница 5: ...us situation that can cause damage to property or mild to moderate injuries WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER indicates a maximal h...

Страница 6: ...ication If the soft tissue in the oral cavity is injured the compressed air may enable sep tic substances to enter the tissue Treatment must be discontinued with instruments operated by compressed air...

Страница 7: ...roducts and care systems described therein must be used KaVo recommends specifying a service interval at the dental office for a licensed shop to clean serv ice and check the functioning of the medica...

Страница 8: ...t national statutory regulations Proper use According to these regulations this medical device may only be used for the de scribed application by a knowledgeable user The following must be observed th...

Страница 9: ...us to start up the medical device after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature o...

Страница 10: ...spray ring on the MULTIflex LUX MULTIflex LED coupling 4 2 Check the amount of water CAUTION Overheating of the tooth due to lack of cooling water Thermal damage to the dental pulp Adjust the water am...

Страница 11: ...rings CAUTION Missing or damaged O rings Malfunctions and premature failure Make sure that all O rings are on the coupling and undamaged Number of available O rings 5 Instructions for use GENTLEsilenc...

Страница 12: ...ufficient spray water can cause the medical device to overheat and damage the tooth Check spray water channels and if necessary clean spray nozzles with the nozzle needle Mat no 0 410 0921 Clean the w...

Страница 13: ...uired accessories Tap water 30 C 5 C 86 F 10 F Brush such as a medium hard toothbrush Brush off under flowing tap water 6 2 2 Automated external cleaning KaVo recommends thermodisinfectors in accordan...

Страница 14: ...infec tors in accordance with EN ISO 15883 1 6 2 4 Automated internal cleaning KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are operated with alkaline cleaning agents wit...

Страница 15: ...agents from the KaVo care system Follow the instructions for use of the disinfectant 6 3 3 Machine disinfection external and internal KaVo recommends thermodisinfectors in accordance with EN ISO 1588...

Страница 16: ...nd malfunctions from improper servicing and care Reduced product life Perform proper care regularly Note KaVo only guarantees that its products will function properly when the care prod ucts used are...

Страница 17: ...and care unit with expansion pressure for effective cleaning and care KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before e...

Страница 18: ...ct Immediately remove the product from the steam steriliser after the sterilisa tion cycle The KaVo medical device has a maximum temperature resistance up to 138 280 4 F Select a suitable procedure de...

Страница 19: ...nsert for turbines 0 411 9902 Cleanpac 10 units 0 411 9691 MULTIflex spray head nozzle 0 411 9921 STERIclave bags 0 411 9912 Material summary Mat no Adaptor INTRAmatic CLEANspray and DRYspray 1 007 17...

Страница 20: ...non compliance with operating maintenance or connection instructions calcination or corrosion contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in...

Страница 21: ......

Страница 22: ...1 003 6890 kb 20101109 02 en...

Отзывы: