Instructions for use EXPERTsurg LUX – 1.008.3500
6 Reprocessing steps in accordance with ISO 17664 | 6.5 Storage
53 / 68
The medical device has a maximum temperature resistance up to 138
o
C
(280.4
o
F).
The following parts of the unit are released for sterilization:
▪ Handpiece holder
▪ Motor and motor cable
Sterilization parameters:
Sterilization of the product is approved for the following “wrapped” steam steril-
ization cycles:
▪ Sterilizer with triple pre-vacuum:
– 3 minutes at 135 °C (275 °F)
Drying time: 16 min.
▪ Sterilizer using the gravity method:
– 10 minutes at 135 °C (275 °F)
Drying time: 30 min.
4
Remove the medical device from the sterilizer immediately after completion
of the sterilization cycle.
4
Use in accordance with the manufacturer's Instructions for Use.
NOTE
Allow the sterilized items to cool to room temperature before re-use.
6.5 Storage
4
Comply with all hygiene requirements when storing sterile items.
4
Store in a dry place protected from dust.
4
Note the labeling on the packaging.
4
Note the duration of storage.