6 Reprocessing steps in accordance with ISO 17664 | 6.1 Manual processing
48 / 68
6 Reprocessing steps in accordance with ISO 17664
NOTE
The reprocessing steps for surgical motors with motor cable and for the
straight and contra-angle handpieces are described in the corresponding In-
structions for use.
NOTE
The processing instructions have been validated by the manufacturer. Any de-
parture from the instructions provided must be checked by the processing en-
tity for efficacy and possible detrimental consequences.
6.1 Manual processing
6.1.1 Manual cleaning
4
Use a soft disposable cloth to wipe down all visible surfaces of the unit, bot-
tle holder, foot control surfaces, and connecting cables.
Manual cleaning of the foot control
4
On the underside of the foot control, press down the snap-in lug
①
of the
speed button
②
and take the speed button
②
off the foot control.
1
2