Instructions for use ESTETICA E30
4 Operation | 4.11 Using instruments
90 / 134
▪ Cannula adapter, reducing handpiece to 7 mm
②
(
Mat. no. 0.764.5873
)
▪ Cannula adapter, reducing handpiece to 11 mm
②
(
Mat. no. 0.764.5883
)
Vacuum stop
CAUTION
Danger of backflow
Swallowing or choking hazard for the patient
▶
Only actuate the vacuum stop when the suction cannula is not in the pa-
tient's mouth.
Note
Vacu-Stop is available in combination with Venturi suction, external wet suc-
tion system, and suction valve.
When the base switch is actuated, the suction of the removed hose is stopped.
①
Vacu-Stop
4.11.3 Using triple-function handpieces
CAUTION
Cannulas that are worn, not screwed in or not locked into place.
Injury from swallowing the cannula.
▶
Before each treatment, ensure that the cannula is properly locked or
screwed into place and firmly seated.
▶
Use original KaVo cannulas only.
CAUTION
Risk of injury from touching the cheek with the handpiece.
Irritation of the mucosa.
▶
Rotate the cannula of the handpiece into an operating position where there
is no contact of the mucosa.