Instructions for use ESTETICA E30
7 Safety checks - testing instructions | 7.2 Instructions for safety checks
116 / 134
①
Assistant element: Fastening screw
on the bottom of the assistant ele-
ment
Scan the treatment lamp with the test tip
Operating light KaVoLUX 540 LED U
①
Fastening screw of the handle sup-
port when the gripping sleeve has
been removed
Operating light EDI/MAIA
No measuring points need to be scanned on the operating lights EDI and MAIA.
Measure protective conductor resistance of accessories
See also:
2
8 Annex - Additional measuring points, Page 122
Measure equivalent unit leakage current
Limit
< 10 mA (maximum value!)