Instructions for use DIAGNOdent pen 2190
5 Operation | 5.4 Menus
Note
The probes must be checked for damage before and after each use. They may only
be used with the DIAGNOdent pen 2190 and only for the probe memory for which
they were calibrated. The probe must not be scratched by scalpels, other probes,
tweezers, etc. Do not drop the probe!
1
2
3
▶ Place the proximal attachment ② on the reference ③ only when calibrating the
proximal probe ①.
ð
Depending on the working direction of the probe, the values can differ slightly for
technical reasons; however, it is generally not necessary to recalibrate during
treatment.
▶ Press the menu key.
ð
The calibration symbol appears.
Note
Do not point the probe toward light sources or reflective surfaces.
▶ Press the save key.
ð
Calibration is started.
▶ Once you hear the signal, place the probe vertically on the reference.
ð
As soon as the signal tone stops, calibration is over.
ð
Calibration is successful when the value is the display agrees with the reference
value (± 3).
23 / 44